Skip to main navigation Skip to search Skip to main content

Laboratory evaluation of a dual-path platform assay for rapid pointof-care HIV and syphilis testing

  • S. R. Leon
  • , L. B. Ramos
  • , S. K. Vargas
  • , N. Kojima
  • , D. G. Perez
  • , C. F. Caceres
  • , J. D. Klausner
  • Universidad Peruana Cayetano Heredia
  • Socios en Salud Lima
  • University of California
  • Hospital Nacional Daniel Alcides Carrion

Research output: Contribution to journalArticlepeer-review

20 Scopus citations

Abstract

We assessed the laboratory performance of the Chembio dual-path platform HIV-syphilis rapid immunodiagnostic test and electronic reader for detection of HIV and Treponema pallidum antibodies in 450 previously characterized serum specimens. For visual or electronic reader HIV antibody detection, the sensitivity was 100% and the specificity was 98.7%. For visual T. pallidum antibody detection, the test sensitivity was 94.7% and the specificity was 100.0%; with the electronic reader, the sensitivity was 94.7% and the specificity was 99.7%.

Original languageEnglish
Pages (from-to)492-494
Number of pages3
JournalJournal of Clinical Microbiology
Volume54
Issue number2
DOIs
StatePublished - Feb 2016

UN SDGs

This output contributes to the following UN Sustainable Development Goals (SDGs)

  1. SDG 3 - Good Health and Well-being
    SDG 3 Good Health and Well-being

Fingerprint

Dive into the research topics of 'Laboratory evaluation of a dual-path platform assay for rapid pointof-care HIV and syphilis testing'. Together they form a unique fingerprint.

Cite this