T1 - Evaluation of the Performance of Portable Hemoglobinometers at Measuring Hemoglobin and Detecting Anemia in a Periurban Pediatric Population in Lima, Peru
AU - Silverstein, Zach
AU - Mantari, Alicia Alva
AU - Breglio, Kimberly
AU - Pajuelo, Monica J.
AU - Valdiviezo, Gladys
AU - Cruz, Sandra
AU - Castañeda, Pamela
AU - Rothstein, Jessica D.
AU - Huaman-Fernandez, Lucia
AU - Cabrera, Lilia
AU - Salguedo, Mario
AU - Barrientos, Franklin
AU - Saldivar-Espinoza, Bryan
AU - Nuóñez-Fernandez, Dennis
AU - Gilman, Robert H.
AU - Sheen, Patricia
AU - Roman-Gonzalez, Avid
AU - Noazin, Sassan
AU - Zimic, Mirko
N1 - Publisher Copyright:
Copyright © 2025 American Society of Tropical Medicine and Hygiene.
PY - 2025/3
Y1 - 2025/3
N2 - As many as one in three people worldwide have anemia, with young children at increased risk of both disease and complications. In settings without clinical laboratories, portable hemoglobinometers serve important roles in diagnosing anemia and estimating prevalence. Here, we assess the validity of two such point-of-care devices—the HemoCue Hb201 and the HemoCue Hb301—relative to the international reference standard, the cyanmethemoglobin method. In total, 428 children ages 6–60 months were recruited at health posts in Lima, Peru, and venous and capillary blood samples were collected from each participant. Venous blood was assessed with the cyanmethemoglobin method, whereas capillary blood was assessed using the Hb201 and the Hb301; 16.1% of participants were found to have anemia using the cyanmethemoglobin method. Both the Hb201 (43.7%) and the Hb301 (20.6%) overestimated this prevalence, with the former reaching statistical significance (P,0.0001 and P 5 0.11, respectively). Both devices also tended to underestimate hemoglobin concentration, with the Hb201 (mean difference 5 20.99 g/dL; percentage error 5 28.1%) being appreciably less accurate than the Hb301 (mean difference 5 20.35 g/dL; percentage error 5 22.7%). Areas under the curve for the Hb201 (0.92) and the Hb301 (0.93) were statistically similar (P 5 0.28); however, the Hb201 incorrectly classified 29.4% of participants compared with 11.0% for the Hb301. Both devices had more false positives than false negatives. In conclusion, the Hb301 was found to be significantly more accurate than the Hb201 at measuring hemoglobin, diagnosing anemia, and estimating anemia prevalence.
AB - As many as one in three people worldwide have anemia, with young children at increased risk of both disease and complications. In settings without clinical laboratories, portable hemoglobinometers serve important roles in diagnosing anemia and estimating prevalence. Here, we assess the validity of two such point-of-care devices—the HemoCue Hb201 and the HemoCue Hb301—relative to the international reference standard, the cyanmethemoglobin method. In total, 428 children ages 6–60 months were recruited at health posts in Lima, Peru, and venous and capillary blood samples were collected from each participant. Venous blood was assessed with the cyanmethemoglobin method, whereas capillary blood was assessed using the Hb201 and the Hb301; 16.1% of participants were found to have anemia using the cyanmethemoglobin method. Both the Hb201 (43.7%) and the Hb301 (20.6%) overestimated this prevalence, with the former reaching statistical significance (P,0.0001 and P 5 0.11, respectively). Both devices also tended to underestimate hemoglobin concentration, with the Hb201 (mean difference 5 20.99 g/dL; percentage error 5 28.1%) being appreciably less accurate than the Hb301 (mean difference 5 20.35 g/dL; percentage error 5 22.7%). Areas under the curve for the Hb201 (0.92) and the Hb301 (0.93) were statistically similar (P 5 0.28); however, the Hb201 incorrectly classified 29.4% of participants compared with 11.0% for the Hb301. Both devices had more false positives than false negatives. In conclusion, the Hb301 was found to be significantly more accurate than the Hb201 at measuring hemoglobin, diagnosing anemia, and estimating anemia prevalence.
UR - https://www.scopus.com/pages/publications/86000498703
U2 - 10.4269/ajtmh.24-0494
DO - 10.4269/ajtmh.24-0494
M3 - Artículo
C2 - 39657230
AN - SCOPUS:86000498703
SN - 0002-9637
VL - 112
SP - 685
EP - 691
JO - American Journal of Tropical Medicine and Hygiene
JF - American Journal of Tropical Medicine and Hygiene
IS - 3
ER -