Non-clinical performance and acceptability of a small portable respiratory stimulator device for basic neonatal resuscitation, tested by health personnel

Título traducido de la contribución: Desempeño no clínico y aceptabilidad de un estimulador respiratorio portátil para reanimación neonatal básica, evaluado en personal de salud
  • Carlos A. Delgado
  • , Víctor M. Sánchez
  • , Pablo M. Velásquez
  • , Roberto L. Shimabuku
  • , Luis Huicho
  • Univ. Nacional Mayor de San Marcos
  • Universidad Peruana Cayetano Heredia
  • Instituto Nacional de Salud Del Niño
  • Instituto Nacional Materno Perinatal, Lima

Producción científica: Contribución a una revistaArtículorevisión exhaustiva

Resumen

Introduction: Neonatal resuscitation demands equipment for respiratory support not always available in rural areas. Innovative devices are required, and rapid prototyping allows to generate them using three-dimensional (3D) designs and printers. Objective: To evaluate the non-clinical performance and the acceptability by health personnel of a neonatal respiratory device produced by rapid prototyping. Methods: Observational study, descriptive, of proof of concept developed in two steps. Step 1: Manufacture of the device with rapid prototyping in three-dimensional (3D) scanners and printers. Step 2: Demonstration of the invention during training programs in neonatal resuscitation for health personnel in three regions of Peru (Tarapoto, Huánuco and Ayacucho). In both steps, we evaluated the performance of the device connected to a gas flow analyser. A survey was administered to the health workers of Tarapoto and Ayacucho to know their acceptability. Results: The developed prototype is T-shaped with two side bellows that, when pressed with one hand, project air through the centre towards a facial adapter. The use of the prototype in the laboratory generated an average air flow of 4.8 Lt /min (SD ± 1.7) and an average pressure of 5.9 cmH2O (SD ± 1.4). This device was considered to be "very simple to use" in an acceptability survey involving 39 nurses and 11 doctors in remote areas of the capital of Peru. Conclusions: The evaluated prototype is acceptable by the staff and has a performance capable of generating spontaneous breathing at birth.

Título traducido de la contribuciónDesempeño no clínico y aceptabilidad de un estimulador respiratorio portátil para reanimación neonatal básica, evaluado en personal de salud
Idioma originalInglés
Páginas (desde-hasta)298-304
Número de páginas7
PublicaciónAnales de la Facultad de Medicina
Volumen80
N.º3
DOI
EstadoPublicada - jul. 2019