Performance of binaxNOW G6PD deficiency point-of-care diagnostic in P. vivax-infected subjects

  • Lyda Osorio
  • , Nick Carter
  • , Preetam Arthur
  • , Germana Bancone
  • , Sowmya Gopalan
  • , Sandeep K. Gupta
  • , Harald Noedl
  • , Sanjay K. Kochar
  • , Dhanpat K. Kochar
  • , Srivicha Krudsood
  • , Marcus V. Lacerda
  • , Alejandro Llanos-Cuentas
  • , Ronnatrai Rueangweerayut
  • , Ramadurai Srinivasan
  • , Moritz Treiber
  • , Jörg J. Möhrle
  • , Justin Green
  • Universidad Del Valle
  • GlaxoSmithKline plc.
  • Sri Ramachandra Institute for Higher Education and Research
  • Shoklo Malaria Research Unit
  • M.V. Hospital and Research Centre
  • Medical University of Vienna
  • Sardar Patel Medical College
  • Rajasthan University of Health Sciences
  • Faculty of Tropical Medicine, Mahidol University
  • Fundação de Medicina Tropical Doutor Heitor Vieira Dourado
  • Mae-Sot General Hospital
  • Clinical Development

Producción científica: Contribución a una revistaArtículorevisión exhaustiva

27 Citas (Scopus)

Resumen

Accurate diagnosis of glucose-6-phosphate dehydrogenase (G6PD) deficiency is required to avoid the risk of acute hemolysis associated with 8-aminoquinoline treatment. The performance of the BinaxNOW G6PD test compared with the quantitative spectrophotometric analysis of G6PD activity was assessed in 356 Plasmodium vivax-infected subjects in Brazil, Peru, Thailand, and India. In the quantitative assay, the median G6PD activity was 8.81 U/g hemoglobin (range = 0.05-20.19), with 11 (3%) subjects identified as deficient. Sensitivity of the BinaxNOW G6PD to detect deficient subjects was 54.5% (6 of 11), and specificity was 100% (345 of 345). Room temperatures inadvertently falling outside the range required to perform the rapid test (18-25°C) together with subtlety of color change and insufficient training could partially explain the low sensitivity found. Ensuring safe use of 8-aminoquinolines depends on additional development of simple, highly sensitive G6PD deficiency diagnostic tests suitable for routine use in malaria-endemic areas.

Idioma originalInglés
Páginas (desde-hasta)22-27
Número de páginas6
PublicaciónAmerican Journal of Tropical Medicine and Hygiene
Volumen92
N.º1
DOI
EstadoPublicada - 1 ene. 2015